Product Specifications 4
Agilent U2802A 31-Channel Thermocouple Input User’s Guide 53
Tabl e 4-3 U2802A measurement accuracy with U2356A, at 0 to 18 °C and 28 to 45 °C,
with different number of averaging points.
Thermocouple Measurements Accuracy
(U2356A @ 0 °C to 18 °C and 28 °C to 45 °C)
T/C
Ty pe
ITS-90
Temperature
Range (°C)
Optimum
Measurement
Range(°C)
Without
averaging
(± °C)
50 points
averaging
(± °C)
500 points
averaging
(± °C)
Low High Low High
B 0 1820 1100 1820 6.1 3.1 2.4
400 1100 14.4 6.3 2.7
E –270 1000 –150 1000 3.0 2.6 2.6
–200 –150 4.2 3.7 3.6
J –210 1200 –150 1200 2.9 2.5 2.5
–210 –150 4.9 4.1 4.0
K –270 1372 –100 1200 3.3 2.9 2.9
–200 –100 5.3 4.2 4.0
N –270 1300 –100 1300 3.4 2.7 2.6
–200 –100 6.8 4.6 4.1
R –50 1768 300 1760 6.2 3.7 3.2
–50 300 15.7 7.2 3.8
S –50 1768 400 1760 6.4 4.0 3.4
–50 400 14.2 6.6 3.4
T –270 400 –100 400 3.0 2.4 2.2
–200 –100 5.3 4.2 3.9
U2802A_UG.book Page 53 Tuesday, January 8, 2008 10:24 AM